Specifications
• Accuracy
ADT672-02: 0.02% of full scale
ADT672-05: 0.05% of full scale
• Gauge Types
Gauge pressure
Compound Pressure
Absolute pressure
Differential pressure
• Display
Description: Dual-line 6 full digit FSTN LCD with LED backlight
Display rate: 3.5 readings per second (Default setting)
Numeral display height: 16.5mm (0.65″)
• Pressure Units
Pa, kPa, MPa, psi, bar, mbar, inH2O 4°C, mmH2O 4°C, inHg 0°C, mmHg 0°C, kgf/cm2
• Environmental
Compensated Temperature: 14°F to 122°F (-10°C to 50°C)
Operating Temperature: 14°F to 122°F (-10°C to 50°C)
Storage Temperature: -4°F to 158°F (-20°C to 70°C)
Humidity: <95%
• Pressure Connections
1/4NPT male, 1/4BSP male, M20X1.5 male (≤15,000 psi)
1/4HP female or 1/4HP male (>15,000 psi)
0.236 inch (Ø6mm) test hose (for differential pressure)
Other connections available per request
• Electrical Connection
0.156 inch (Ø4mm) sockets
• Over Pressure Warning:
120%
• Data Logger
Storage capacity: 30 files, 40 records per file
Mode: manual and automatic
Hourly-record: record the data every hour
Interval-record: set by user
• Power
Battery: Rechargeable Li-ion polymer battery
Li-Battery working time: 40 hours
Recharge time: 4 hours
External power: 110V/220V power adapter (DC10V)
• Enclosure
Case material: Aluminum alloy
Wetted parts: 316L SS
Dimension: Ø120mm X 46mm depth X 184mm height
Weight: 0.7kg
European Compliance: CE marked
• Communication
RS232 (DB9/F, environmentally sealed)
Baud rate: 1200,2400,4800,9600
Data length: 8 bits
Stop bit: 2 bits
Address: from 1 to 112
• Warranty:
1 year
• Pressure Ranges
Gauge Pressure [1]
|
P/N
|
Pressure Range
|
Media[2]
|
Accuracy(%FS)
|
Burst
Pressure
|
(psi)
|
(bar)
|
V15
|
-15
|
-1.0
|
G
|
0.02 (0.05)
|
3×
|
GP2
|
2
|
0.16
|
G
|
0.05
|
3×
|
GP5
|
5
|
0.35
|
G
|
0.05
|
3×
|
GP10
|
10
|
0.7
|
G
|
0.05
|
3×
|
GP15
|
15
|
1.0
|
G, L[3]
|
0.02 (0.05)
|
3×
|
GP30
|
30
|
2.0
|
G, L[3]
|
0.02 (0.05)
|
3×
|
GP50
|
50
|
3.5
|
G, L
|
0.02 (0.05)
|
3×
|
GP100
|
100
|
7.0
|
G, L
|
0.02 (0.05)
|
3×
|
GP300
|
300
|
20
|
G, L
|
0.02 (0.05)
|
3×
|
GP500
|
500
|
35
|
G, L
|
0.02 (0.05)
|
3×
|
GP600
|
600
|
40
|
G, L
|
0.02 (0.05)
|
3×
|
GP1K
|
1,000
|
70
|
G, L
|
0.02 (0.05)
|
2×
|
GP2K
|
2,000
|
140
|
G, L
|
0.02 (0.05)
|
3×
|
GP3K
|
3,000
|
200
|
G, L
|
0.02 (0.05)
|
3×
|
GP5K
|
5,000
|
350
|
G, L
|
0.02 (0.05)
|
3×
|
GP10K
|
10,000
|
700
|
G, L
|
0.02 (0.05)
|
2×
|
GP15K
|
15,000
|
1,000
|
G, L
|
0.05(0.1)
|
2×
|
GP20K
|
20,000
|
1,400
|
G, L
|
0.05(0.1)
|
1.5×
|
GP25K
|
25,000
|
1,600
|
G, L
|
0.1
|
1.5×
|
GP30K
|
30,000
|
2,000
|
G, L
|
0.1
|
1.5×
|
GP36K
|
36,000
|
2,500
|
G, L
|
0.1
|
1.5×
|
GP40K
|
40,000
|
2,800
|
G, L
|
0.1
|
1.1×
|
Note: [1]. Sealed gauge pressure for above 1000 psi
[2]. G=Gas, L=Liquid (please specify media type when place order)
[3]. 0.025% FS for gas media only
Compound Pressure
|
P/N
|
Pressure Range
|
Media
|
Accuracy(%FS)
|
Burst
Pressure
|
(psi)
|
(bar)
|
CP2
|
±2
|
±0.16
|
G
|
0.05
|
3×
|
CP5
|
±5
|
±0.35
|
G
|
0.05
|
3×
|
CP10
|
±10
|
±0.7
|
G
|
0.05
|
3×
|
CP15
|
±15
|
±1
|
G
|
0.02 (0.05)
|
3×
|
CP30
|
-15 to 30
|
-1 to 2
|
G
|
0.02 (0.05)
|
3×
|
CP100
|
-15 to 100
|
-1 to 7
|
G, L
|
0.02 (0.05)
|
3×
|
CP300
|
-15 to 300
|
-1 to 20
|
G, L
|
0.02 (0.05)
|
3×
|
Absolute Pressure
|
P/N
|
Pressure Range
|
Media[2]
|
Accuracy(%FS)
|
Burst
Pressure
|
(psi)
|
(bar)
|
AP5
|
5
|
0.35
|
G
|
0.1
|
3×
|
AP10
|
10
|
0.7
|
G
|
0.1
|
3×
|
AP15
|
15
|
1.0
|
G
|
0.1
|
3×
|
AP30
|
30
|
2.0
|
G
|
0.1
|
3×
|
AP50
|
50
|
3.5
|
G
|
0.1
|
3×
|
AP100
|
100
|
7.0
|
G , L
|
0.05 (0.1)
|
3×
|
AP300
|
300
|
20
|
G , L
|
0.05 (0.1)
|
3×
|
AP500
|
500
|
35
|
G , L
|
0.05 (0.1)
|
3×
|
AP1K
|
1,000
|
70
|
G , L
|
0.05 (0.1)
|
2×
|
AP3K
|
3,000
|
200
|
G , L
|
0.05 (0.1)
|
3×
|
AP5K
|
5,000
|
350
|
G , L
|
0.05 (0.1)
|
3×
|
Differential Pressure
|
P/N
|
Pressure Range
|
Media
|
Accuracy(%FS)
|
Burst
Pressure
|
(inH2O)
|
(mbar)
|
DP1
|
±1
|
±2.5
|
G
|
0.05
|
100×
|
DP2
|
±2
|
±5.0
|
G
|
0.05
|
100×
|
DP5
|
±5
|
±10
|
G
|
0.05
|
50×
|
DP10
|
±10
|
±25
|
G
|
0.05
|
20×
|
DP20
|
±20
|
±50
|
G
|
0.05
|
20×
|
DP30
|
±30
|
±75
|
G
|
0.05
|
15×
|
DP50
|
±50
|
±160
|
G
|
0.05
|
3×
|
DP150
|
±150
|
±350
|
G
|
0.02
|
3×
|
DP300
|
±300
|
±700
|
G
|
0.02
|
3×
|
* One year accuracy (including yearly stability) except DP1 and DP2 range which is 0.05%FS calibration accuracy and 0.05%FS yearly stability.
Electrical Measurement and Source Accuracy
|
|
Range
|
Accuracy
|
Voltage DC
|
±30.0000V
|
± (0.01%RD+1.5mV)
|
Current DC
|
±30.0000mA
|
± (0.01%RD+1.5μA)
|
DC 24V
|
24V±0.5V max:50mA Protect at: 120mA
|
Switch[1]
|
Status OPEN/CLOSED
|
Note: [1] 1V~12V if switch has detective voltage